• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 10CM; SCS EXTENSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EXTENSION, DUAL 4 CHANNEL 10CM; SCS EXTENSION Back to Search Results
Model Number 3341
Device Problems Disconnection (1171); Impedance Problem (2950)
Patient Problem Inadequate Pain Relief (2388)
Event Date 07/24/2018
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 3.Reference mfr.Report#1627487-2018-07839, reference mfr.Report#3006705815-2018-01988.It was reported post-operative implant reprogramming, the patient's scs extension was detached from the ipg header confirmed by x-rays.Reportedly, diagnostic testing revealed invalid impedances on multiple contacts.As a result, an additional surgery was undertaken wherein the extension was reinserted back into the header which resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION, DUAL 4 CHANNEL 10CM
Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
melissa nkematu
6901 preston road
plano, TX 75024
9723092520
MDR Report Key7807759
MDR Text Key117955090
Report Number1627487-2018-07838
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05414734402279
UDI-Public05414734402279
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/18/2020
Device Model Number3341
Device Lot Number6281771
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/24/2018
Initial Date FDA Received08/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/18/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age25 YR
-
-