• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number RNS-300M-K
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Discomfort (2330)
Event Date 07/27/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
The patient reported pain and discomfort at implant site.To improve patient comfort, the decision was made to return to the operating room where the skull was shaved around the ferrule to allow it to sit more flush with the skull.A new ferrule was then implanted and the existing neurostimulator placed back in the ferrule.There is no indication of product malfunction or failure.The rns neurostimulator remains implanted and programmed for detection and stimulation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key7807849
MDR Text Key117959671
Report Number3004426659-2018-00029
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005120
UDI-Public010085554700512017171013
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model NumberRNS-300M-K
Device Catalogue Number1007603
Device Lot Number23567-1-1-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/27/2018
Initial Date FDA Received08/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age36 YR
-
-