The customer stated that they received an erroneous result for one patient sample tested for crep2 creatinine plus ver.2 on the cobas 8000 c 702 module.The sample initially resulted with a crep2 value of 1.47 mg/dl, which was reported outside of the laboratory.The sample was repeated on (b)(6) 2018 and it resulted as 1.72 mg/dl.The sample was repeated on an i-stat analyzer using a biosensor manufactured by abbott resulting as 5.4 mg/dl.The 5.4 mg/dl value was reported outside of the laboratory to the doctor.No adverse events were alleged to have occurred with the patient.The patient's status is good.The serial number of the c 702 analyzer is (b)(4).
|