Device 2 of 3.Reference mfr.Report#: 1627487-2018-07849.Reference mfr.Report#: 1627487-2018-07851.It was reported during an ipg replacement procedure on (b)(6) 2018 (reference mfr.Report#: 1627487-2018-0763), it was found the leads had pulled out of the ipg header.As such, the leads were reconnected to the new ipg.Intra-op testing revealed high and low impedance values on multiple lead contacts.However, the company representative used the valid lead contacts and was able to reprogram and provide the patient with effective stimulation therapy.
|