Catalog Number 07D63-21
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
A product recall letter was issued to all architect customers who have received architect creatine kinase, list number 07d63-21, lot number 87697un18.The letter informs the customer of the issue regarding a stability issue that may lead to quality control results shifting low out of range and/or error code 1054 "unable to calculate result, reaction check failure" for quality control and patient samples.The letter instructs the customer to discontinue use of the suspect lots and destroy any remaining inventory.The cause of the stability failure has been traced to a contamination of the bulk material.
 
Event Description
The customer observed error code 1054 unable to calculate result, reaction check failure using the architect creatine kinase assay (list number 7d63, lot number 87697un18).No impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT CREATINE KINASE
Type of Device
CREATINE KINASE (CK)
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer (Section G)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key7816491
MDR Text Key118246307
Report Number1628664-2018-01626
Device Sequence Number1
Product Code CGS
UDI-Device Identifier00380740012878
UDI-Public00380740012878
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K983070
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 08/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/22/2018
Device Catalogue Number07D63-21
Device Lot Number87697UN18
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction Number1628664-07/19/18-002-R
Patient Sequence Number1
Treatment
(B)(4); ARCHITECT C8000, LIST 01G06,
-
-