Device Problem Migration or Expulsion of Device (1395)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 06/18/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that a full term new born received an endeavor resolute stent deployed in a straight short conical duct.On the second post-natal day the stent migrated, possibly due to inadequate ductal constriction, persisting prostaglandin effect, and straight short ductal morphology.It was successfully retrieved using a larger balloon and redeployed in the duct again.Intravenous indomethacin was administered to prevent further stent migration.The patient was discharged after 3 days.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDEAVOR RESOLUTE RX
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key7818507
MDR Text Key118291810
Report Number9612164-2018-02185
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P110013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/15/2018
Initial Date FDA Received08/27/2018
Supplement Dates Manufacturer Received01/22/2019
Supplement Dates FDA Received01/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age2 DA
Patient Weight2
-
-