MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA¿ STOPCOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA¿ STOPCOCK Back to Search Results
Catalog Number 394995
Device Problem Disconnection (1171)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/06/2018
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone#: (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that during use a bd connecta¿ stopcock "the extension disengages from the faucet" no report of serious injury or medical intervention.
 
Manufacturer Narrative
The device history report for lot number 8037787 was reviewed, no related abnormalities were found.Material number (b)(4) for lot 8037787 was manufactured on feb-18-2018.This is the first instance of this failure mode with this lot number.According to the sampling plan applied for product performance, this lot was accepted and released, with no defects being noted during final assembly or in packaging visual inspections.Unfortunately a sample could not be obtained for the purpose of the investigation.Investigation conclusion: without the ability to investigate the affected unit our quality engineers were unable to determine the root cause for this complaint.Based on investigation results to date, root cause for manufacturing process cannot be determined.
 
Event Description
It was reported that during use a bd connecta stopcock "the extension disengages from the faucet" no report of serious injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CONNECTA¿ STOPCOCK
Type of Device
STOPCOCK
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7818795
MDR Text Key118529528
Report Number9610847-2018-00272
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date01/31/2021
Device Catalogue Number394995
Device Lot Number8037787
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/11/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-