Catalog Number 66800041
Device Problem Obstruction of Flow (2423)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/30/2018
Event Type  malfunction  
Event Description
It was reported that during debridement utilizing versajet hand-piece, the saline jet stopped.The surgery was completed with surgical knives.There was no patient harm.
 
Manufacturer Narrative
We have now completed our investigation into the returned versajet hand piece.A visual inspection was carried out and no problems were found.Functional testing was carried out and this confirmed the reported issue.The device was found to leak water and the location of the leak was at the weld where the jet tube meets the filter assembly.Unfortunately on this occasion we were unable to determine the root cause of this failure and a device history review was not possible as no lot number was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAJET EXACT ASSY, 45 DEGREE X 14MM
Type of Device
LAVAGE, JET
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD
101 hessle road
hull, HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD
101 hessle road
hull, HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
sarah freestone
101 hessle road
hull, HU3 2-BN
UK   HU3 2BN
MDR Report Key7819565
MDR Text Key118395054
Report Number8043484-2018-00255
Device Sequence Number1
Product Code FQH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K143115
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66800041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/31/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-