ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 19AGFN-756
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/16/2018
Event Type  Injury  
Manufacturer Narrative
The reported event of a dislodged leaflet was confirmed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the dislodgement remains unknown; however it is consistent with an external force applied to the leaflet and orifice, which overstressed the carbon material and resulted in the dislodgement.Please note, per the instructions for use (b)(4) , "to avoid structural damage, the valve must be rotated in the fully closed position.".
 
Event Description
On (b)(6) 2018, a 19mm regent valve was implanted, but when the physician rotated the valve with the abbott holder handle model 905, a leaflet was dislodged.The valve and dislodged leaflet were removed from the patient and another 19mm regent valve was implanted.The patient did not have any consequences.Per report, the procedure was delayed by about half an hour, but the patient was hemodynamically stable throughout the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer Contact
elizabeth boltz
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7820530
MDR Text Key118386686
Report Number2648612-2018-00063
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005852
UDI-Public05414734005852
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/22/2021
Device Model Number19AGFN-756
Device Catalogue Number19AGFN-756
Device Lot Number5746555
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-