ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL
|
Back to Search Results |
|
Model Number 19AGFN-756 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/16/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
The reported event of a dislodged leaflet was confirmed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the dislodgement remains unknown; however it is consistent with an external force applied to the leaflet and orifice, which overstressed the carbon material and resulted in the dislodgement.Please note, per the instructions for use (b)(4) , "to avoid structural damage, the valve must be rotated in the fully closed position.".
|
|
Event Description
|
On (b)(6) 2018, a 19mm regent valve was implanted, but when the physician rotated the valve with the abbott holder handle model 905, a leaflet was dislodged.The valve and dislodged leaflet were removed from the patient and another 19mm regent valve was implanted.The patient did not have any consequences.Per report, the procedure was delayed by about half an hour, but the patient was hemodynamically stable throughout the procedure.
|
|