Event summary: the patient data files showed at least three applications were performed with catheter 2af284 / 49772 on the date of the event with no system notice or any issues.Patient data files also showed at least eight more applications were performed with catheter 2af284 / 59973 with no system notices or issues.Upon visual inspection of catheter, 2af284 / 49772, results showed that the catheter was intact with no apparent issue.Smart chip verification showed that the catheter has been used for three injections.The catheter failed the performance test due to system notice of ((b)(4)) indicating that the refrigerant delivery path was obstructed.The catheter passed the deflection test as per specification.Dissection showed that guide wire lumen kinked on 1.4140 inches from the tip of the catheter.The catheter failed the performance test due to kink on guide wire lumen.In conclusion, the reported (balloon catheter guidewire bent/kink) issue has been confirmed through testing but not confirmed through the data analysis.The balloon catheter failed the returned product inspection due to guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
|
It was reported that prior to cryo ablation procedure, during set up, after deflating the balloon, it was observed that the tip of the balloon catheter was bent.The balloon catheter was then replaced.The procedure continued and the case was completed with cryo.No patient complications have been reported as a result of this event.
|