• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1550400-12
Device Problem Unintended System Motion (1430)
Patient Problems Intimal Dissection (1333); Foreign Body In Patient (2687)
Event Date 06/28/2018
Event Type  Death  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The stent remains in the vessel; the customer reported the delivery system was discarded.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that during a procedure of the proximal circumflex artery, after pre-dilatation, the 4.0 x 12 mm xience sierra stent system was advanced to the lesion.During stent deployment, the stent watermelon seeded and moved into the left main artery, where it was implanted, but blocked the left anterior descending (lad) artery.The lad was wired and ballooned, which resulted in a left main dissection.The patient was rushed to the or for treatment, but expired.No additional information was provided.
 
Manufacturer Narrative
Internal file number - (b)(4).The device was not returned for analysis.A cine was received and reviewed by an abbott vascular clinical specialist.The reviewer concluded that the cine confirms mis-location of stent; death indirectly a result of migration.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The reported patient effects of intimal dissection and death are listed in the xience sierra everolimus eluting coronary stent system instructions for use as known patient effects of coronary stenting procedures.The investigation was unable to determine a conclusive cause for the reported difficult to deploy.A conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key7824904
MDR Text Key118521240
Report Number2024168-2018-06704
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2019
Device Catalogue Number1550400-12
Device Lot Number8042141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/13/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age47 YR
Patient Weight107
-
-