The investigation was carried out based on the available information and the analysis of the logfile.It was found, that the device in question was not in operation or connected to a patient on the date of event, neither on july 27th (as initially reported) nor on july 26th (as reported additionally during investigation).In the course of complaint investigation, the user confirmed that the reported event occurred on (b)(6) 2018 at 06:30am during device check i.E.Without patient involvement.Based on the investigation of the logfile, for the whole period covered by the log ((b)(6) 2018), no indications for a ventilator failure during operation were found.On july 26th, there were interactions during device check, which led to a non-availability of the device, indicating a wrong assembly of breathing hoses resulting in no o2 delivery measured at the y-piece.Since sw 2.0x, an optional test can be configured by the user to be part of the self-test of the device.With this test, the potential misalignment of the patient circuit was detected as specified during self-test prior to use.However, there were no indications for a device failure found around the reported date of event.In case of a ventilator failure during operation, automatic ventilation is no longer available.Manual respectively spontaneous ventilation remains possible.If the ventilator fails during self-test, as it was reported additionally for this specific case, the failure will be detected by the device and a message with interactions is displayed with dedicated fault/cause/remedy information.If the test cannot be repeated successfully, the device is not operational.Finally, based on the investigation, no indications for a device failure or a ventilator failure were found.Only a potential misalignment of the patient circuit could be found and this was detected during self-test prior to use.This misalignment led to a non-availability of the device ¿ if the self-test cannot be repeated successfully, the device is not operational.It is possible that this non-availability of the device was interpreted as ventilator failure.
|