On (b)(6) 2018, abbott point of care (apoc) was contacted by a customer who reported visible smoke coming from i-stat1 analyzer sn (b)(4) after replacing the batteries.The analyzer was replaced at no charge and returning for investigation.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.However, the customer states that 9volt batteries with the red-fused battery carrier was being used at the time of the event.Therefore the analyzer is unlikely to become hot to touch.The product was replaced and returned for investigation.Based on the information available, there were no patient or user related injuries associated with this complaint.
|
Apoc incident #: (b)(4).The investigation was completed on 10/10/2018.Analyzer s/n (b)(4) was returned because it emitted smoke.Failure analysis determined that the cause of the problem was the failure of a 100 f tantalum capacitor (apoc p/n: 013827-01 01) c26 on the main board of the analyzer, between the primary battery and ground.The capacitor had failed such that it created a short circuit between the battery and ground; current flowing through the shorted capacitor caused heat damage, which gave the capacitor a burned appearance.The rechargeable battery in use at the time of the incident was not returned with the analyzer.
|