Product event summary: the device and data files were returned and analyzed.Data files showed that 9 applications were performed with a balloon catheter on the date of event.Visual inspection of the catheter showed that the device was intact with no apparent issues.Smart chip verification indicated that the catheter was used for 1 injection.Insertion and retraction tests could not be performed.The test mapping catheter was stuck inside the shaft under the balloons segment.Inflation showed a kink on guide wire lumen under the balloon.A dissection showed that the guide wire lumen kinked 1.38 inches from the tip inside the balloons.The catheter passed the performance test.In conclusion, the catheter failed the return the inspection due to a guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
|
It was reported that during a cryo ablation procedure, as the mapping catheter was attempted to advance through the balloon catheter, resistance was noted.The balloon catheter was removed, and another attempt was made outside of the patient body without resolve.The balloon catheter was then replaced and the case was completed with cryo.The balloon catheter was returned to the manufacturer, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
|