• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION SWIFT-LOCK ANCHOR; SCS ANCHOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION SWIFT-LOCK ANCHOR; SCS ANCHOR Back to Search Results
Model Number 1192
Device Problem Break (1069)
Patient Problem Discomfort (2330)
Event Type  Injury  
Manufacturer Narrative
The therapy date for the following device is unknown: model: 3662; scs ipg.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 2 of 2.Reference mfr.Report: 1627487-2018-08206.It was reported (b)(6) the patient experienced discomfort at the lead and anchor site.In addition, the anchor was superficial.The lead had reportedly migrated.Surgical intervention was undertaken during which the anchor was observed to be broken.The lead and anchor were explanted and replaced.Stimulation was restored and the discomfort issue was resolved following the procedure.
 
Event Description
Device 2 of 2: reference mfr.Report: 1627487-2018-08206.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIFT-LOCK ANCHOR
Type of Device
SCS ANCHOR
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
palka sharma
9725264823
MDR Report Key7833510
MDR Text Key118783329
Report Number1627487-2018-07730
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067024084
UDI-Public05415067024084
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K092371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/20/2020
Device Model Number1192
Device Lot Number6304339
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-