Model Number N/A
Device Problem Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Date 08/03/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) unit was reviewed and no non-conformances related to the reported event were noted.The fse order the missing screws and put the screws on the helium regulator assembly.Completed installation of the cardiosave and performed functional and safety tests.The iabp passed all the testes.The iabp was returned to the customer for clinical use.
 
Event Description
A getinge field service engineer (fse) discovered that two screws on helium regulators are missing during the installation of the cardiosave intra-aortic balloon pump (iabp).There was no patient involvement, therefore no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7834267
MDR Text Key118894139
Report Number2249723-2018-01501
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/03/2018
Initial Date FDA Received08/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-