Model Number 3286
Device Problems Break (1069); High impedance (1291)
Patient Problems Fall (1848); Inadequate Pain Relief (2388)
Event Date 02/09/2018
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient experienced ineffective stimulation.The patient fell approximately a month ago.Reprogramming was attempted; however patient felt stimulation only in certain positions.Diagnostics indicated high impedances.X-rays indicated a possible break in the lead.Subsequently, surgical intervention was undertaken on (b)(6) 2018 during which the cervical lead was explanted and replaced.Stimulation was restored following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMITRODE S-8 LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
palka sharma
9725264823
MDR Report Key7839153
MDR Text Key118959419
Report Number1627487-2018-08021
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2012
Device Model Number3286
Device Lot Number3162647
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/17/2018
Initial Date FDA Received08/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/14/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 1192, SCS ANCHOR; MODEL 3662, SCS IPG
Patient Outcome(s) Other;
-
-