• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ULTRA FAST-FIX ASSEMBLY - CURVED; STAPLE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ULTRA FAST-FIX ASSEMBLY - CURVED; STAPLE, FIXATION, BONE Back to Search Results
Catalog Number 72201491
Device Problem Device Fell (4014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/14/2018
Event Type  malfunction  
Event Description
It was reported that after t1 and t2 implantation, t2 fell out.A backup device was available to complete the procedure with no significant delay and no patient injuries.
 
Manufacturer Narrative
One curved ultra fast-fix assembly was returned for evaluation.Visual assessment showed the depth limiter has been trimmed to the surgeons desired length.The trigger is fully advanced and locked within the handle.No ts or suture remain on the needle, however a 12 inch length of suture was returned.Examination of the suture showed no abnormalities.This investigation could not identify any evidence of product contribution to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA FAST-FIX ASSEMBLY - CURVED
Type of Device
STAPLE, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key7841834
MDR Text Key119328324
Report Number1219602-2018-01169
Device Sequence Number1
Product Code JDR
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K972326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/08/2019
Device Catalogue Number72201491
Device Lot Number50518954
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2018
Date Manufacturer Received09/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age45 YR
-
-