The device has not been returned to olympus for evaluation.The cause of the reported complaint cannot be confirmed.Despite multiple followup attempts by phone and in writing, no further information is available regarding the role or performance of the forceps device in the event, the name of the procedure, the sequence of events, whether and how the procedure was completed, the patient¿s current state, or the make and model number of the stone crushing device.It has also not been confirmed if the forceps device is the same device as the stone crusher.The oem performed a review of the device history record and found no related anomalies at the time of manufacturing release.As a preventive measure, the forceps instructions for use document has pre-procedure instructions to visually inspect the product and verify manual operation.The document also states, ¿if the instrument is damaged or does not function properly, replace it.¿ the document also specifies the compatible olympus devices for use with the forceps, and states that for maintenance, the distal tip should be lubricated with olympus a0273 lubricant.
|
Two of 2.Olympus was informed that during an unknown procedure, the forceps device was present when a separate cystoscope sheath device became bent while inside the patient, trapping and breaking a stone crusher.All device fragments were retrieved from the patient.There was no reported patient injury.
|