Model Number DTBA1D1
Device Problems High impedance (1291); Mechanical Problem (1384); Decreased Sensitivity (2534); Impedance Problem (2950)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/08/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that one month after the device changeout, there was a lead integrity alert for high and undefined pacing impedances on the right ventricular lead that was fluctuating.The pocket was reopened to check for a setscrew issue.When checking the rv lead through the analyzer, satisfactory parameters were obtained.After connecting to the device and placing it back into the pocket, there was increasing impedance and loss of rv sensing.There was improvement in impedance and sensing on slight movement of the device.A possible setscrew issue was considered and multiple attempts were made to screw the lead into the device.The device remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA XT
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7844150
MDR Text Key119117790
Report Number3004209178-2018-19853
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00643169530140
UDI-Public00643169530140
Combination Product (y/n)N
Reporter Country CodeCJ
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/14/2018
Device Model NumberDTBA1D1
Device Catalogue NumberDTBA1D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2018
Initial Date FDA Received09/04/2018
Date Device Manufactured04/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight98
-
-