One 72201993 osteoraptor 2.3mm suture anchor returned.The complaint stated: "it was reported that during a glenoid labrum repair surgery, the anchor of osteoraptor broke." the insertion shaft device, suture and partial anchor were returned.There is no damage, bow or bending of the insertion shaft.The ultrabraid suture was returned attached to the handle.The anchor was attached to suture through the eyelet but no longer attached to the distal tip of the inserter.There are proximal barbs missing.Material missing where it attaches to the inserter.The anchor has been fractured at the location where the inserter seats inside the anchor.This condition is consistent with force and potentially losing axial alignment.A starter instrument is mandatory for this device's prep.This product recommends and specified drill bit, a 2.3mm in line drill and a 2.3mm obturator.These are purchased separately.No root cause related to the manufacturing of this device was established.
|