(b)(4).Device evaluated by manufacturer: the device was received with the stent partially deployed from the device.An attempted to deploy the stent was made but was unsuccessful due to a complete break of the outer shaft located approximately 31mm distal of the main valve and a severe kink on the shaft located approximately 15mm distal of the guidewire port.This type of damage is consistent with excessive force being applied to the device when attempting to deploy the stent.A visual and tactile examination identified no issues with the tip of the device.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
Reportable based on the product analysis completed on august 10, 2018.It was reported the stent was unable to deploy.The 75% stenosed, target lesion was located in a highly tortuous arotic arch in the left "aci", a non bsc sheath and a 0.14 non bsc guidewire were placed.The stent was prepped with no difficulties and advanced over the wire, through the sheath, to the lesion.The stent was secured in place and then released.It could only be released a third of the stent and it couldn't be advanced or repositioned.The physician advanced the sheath over the entire length and then pulled the stent back into the sheath without further issues.The procedure was completed with a 9.5 carotid wallstent and everything went perfect.The patient was reported to be fine.However, the returned product analysis revealed that the stent was partially deployed.It was further reported the stent was unable to be re-sheathed and was removed from the patient with a third of the stent still partially deployed.
|