MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED) Back to Search Results
Model Number DM0010FAA
Device Problem Device Remains Activated (1525)
Patient Problem Perforation (2001)
Event Type  Injury  
Manufacturer Narrative
Report inclusive.No evaluation was performed as the device was not returned.This dm0010faa perforator with lot number 101/17 was m anufactured by micromar.Device history record was reviewed by micromar and per their report, all the processes were executed according to what was planned and validated.There were no observations or nonconformities during the manufacturing process.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to monitor this complaint type for trends.Physicians:(b)(6).If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that surgeon was experiencing that the cranial perforator easy drill did not stop and damaged the dura.On follow up it was confirmed this occured back in (b)(6) 2018 and was not reported to the device manufacturer.It was also noted that this may have occurred with two different surgeons at this hospital but this has not been confirmed.It was reported that the current condition of the patient is stable.
 
Manufacturer Narrative
The device has been returned and is still pending their evaluation results.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Report not confirmed.Evaluation could not reproduce the reported malfunction of continues to run.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYDRILL PERFORATOR
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)
Manufacturer (Section D)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -390
BR  09980-390
Manufacturer (Section G)
MICROMAR INDUSTRIA E COMERCIO LTDA
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980 -390
BR   09980-390
Manufacturer Contact
natalia matos
cnpj: 53.168.142/0001-29
av. marginal ao corrego da ser
sao paulo,sp 09980--390
BR   09980-390
140575722
MDR Report Key7857510
MDR Text Key119644157
Report Number1625507-2018-00050
Device Sequence Number1
Product Code HBF
UDI-Device Identifier07898959543081
UDI-Public07898959543081
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K141455
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/02/2019
Device Model NumberDM0010FAA
Device Catalogue NumberDM0010FAA
Device Lot Number107/17
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/05/2018
Initial Date FDA Received09/07/2018
Supplement Dates Manufacturer Received09/18/2018
10/25/2018
Supplement Dates FDA Received10/15/2018
10/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-