Device 1 of 4: reference mfr.Report#: 3006705815-2018-02210, reference mfr.Report#: 1627487-2018-08518, reference mfr.Report#: 1627487-2018-08519.It was reported low impedance was found.X-rays were taken and revealed lead migration.In turn, the patient underwent surgical intervention.During the procedure, the patient's anchors were found to be broken.As a result, the doctor decided to explant and replace the patient's leads and anchors.The doctor also electively explanted and replaced the patient's ipg.
|