Event summary: the patient data files showed at least eleven applications were performed with catheter 2af283 / 90103 without any is sues or system notices on the date of event.Visual inspection of the balloon catheter showed the device was intact with no apparent issues.Smart chip verification indicated the catheter was used for eleven injections.The balloon catheter passed the performance test, but inflation showed kink on the guidewire lumen under the balloon segment.Dissection showed a guidewire lumen kink at 1.37 inches from the tip inside the balloons.In conclusion, the reported kink issue was confirmed through testing.The balloon catheter failed the returned product inspection due to a guide wire lumen kink.If information is provided in the future, a supplemental report will be issued.
|