SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION HOSPITAL CART Back to Search Results
Catalog Number 397003-001
Device Problem No Display/Image (1183)
Patient Problems No Patient Involvement (2645); Patient Problem/Medical Problem (2688)
Event Date 08/29/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the issue was observed when the companion hospital cart was not supporting a patient.The companion hospital cart will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The syncardia companion hospital cart is a large cart with wheels into which the syncardia companion 2 driver docks.It is intended for use in the hospital during the temporary total artificial heart (tah-t) implant procedure and subsequent recovery.The companion hospital cart was not supporting a patient.The customer, a syncardia certified hospital, reported that the display on the hospital cart was blank.
 
Manufacturer Narrative
The hospital cart passed functional testing, but additional testing was able to reproduce the customer-reported issue of the screen going blank.The investigation determined the root cause to be a malfunction of the lcd display module likely caused by the lcd printed circuit board (pcb) inverter not having the insulator behind it.This is a known issue that has been previously documented and investigated.All hospital carts returned to syncardia are evaluated and the lcd displays without the insulator are replaced during the hospital cart service procedure.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION HOSPITAL CART
Type of Device
HOSPITAL CART
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7862172
MDR Text Key119808586
Report Number3003761017-2018-00458
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 10/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number397003-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2018
Initial Date Manufacturer Received 08/29/2018
Initial Date FDA Received09/10/2018
Supplement Dates Manufacturer Received08/29/2018
Supplement Dates FDA Received10/26/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-