• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Disconnection (1171); High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Date 07/24/2018
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
Device 2 of 3.Reference mfr.Report#1627487-2018-08609, reference mfr.Report#3006705815-2018-02230.It was reported diagnostic testing identified high impedance values on all contacts during a reprogramming appointment.In turn, x-rays revealed the lead appeared to be pulled out of the header.Reportedly, reprogramming was not attempted due to the issue.As a result, surgical intervention is pending to address the issue.
 
Event Description
Device 2 of 3.Reference mfr.Report#1627487-2018-08609.Reference mfr.Report#3006705815-2018-02230.Follow up revealed that the patient's scs system was explanted and replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
melissa nkematu
6901 preston road
plano, TX 75024
9723092520
MDR Report Key7865657
MDR Text Key119860877
Report Number3006705815-2018-02231
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067017246
UDI-Public05415067017246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/29/2019
Device Model Number3186
Device Lot NumberA000044823
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/22/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/29/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 1192(2), SCS ANCHOR
Patient Outcome(s) Other;
Patient Weight61
-
-