No product samples, cts, lateral radiographs, initial or serial radiographs or other medical records have been provided.Alleged osteolysis at the proximal end of the prosthetic shaft.There is osteolytic resorptive changes at the distal tip of the femoral prosthesis.A medical advisor review of the reporting surgeons notes, cement did not extend beyond the tip of the prosthesis, cortical thinning at the distal tip is likely insignificant without further imaging to confirm.Within in medical probability, at 15 years, the synplug did not cause or contribute to the proximal loosening.
|
Initial hip replacement (hip arthroplasty) surgery involving a hip prosthesis, thp cement, and synplug cement restrictor, left side occurred in 2003.Ct scan was reportedly performed and allegedly a fine lysis margin with pathologically metabolism formed at the proximal portion of the cemented femoral shaft with signs of loosening.The surgeon additionally noted osteolysis in the left distal tip of the prosthetic shaft.Ct scans were not provided.No plans for revision.Surgeon will monitor.
|