Model Number 9-PFO-030
Device Problem Material Deformation (2976)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/29/2018
Event Type  Injury  
Event Description
On (b)(6) 2018, a 30mm amplatzer pfo occluder was prepared and selected according to the ifu using tee and fluoroscopy.The device was deployed and required no repositioning, however, just prior to release, the user observed the discs had deployed deformed.The physician re-sheathed the device, positioned and re-deployed the device with the same result.The device was removed from the patient and a new 30mm amplatzer pfo occluder (lot number: 6457527) was successfully implanted.The patient is reported to be stable.
 
Manufacturer Narrative
The reported event of deformity upon deployment could not be confirmed.The investigation confirmed the device met functional specifications when analyzed at abbott under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PFO OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7879130
MDR Text Key120381874
Report Number2135147-2018-00152
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date11/30/2022
Device Model Number9-PFO-030
Device Catalogue Number9-PFO-030
Device Lot Number6234047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-