On (b)(6) 2018, a 22mm amplatzer septal occluder was selected for use with a 10fr delivery catheter.Upon deployment, the device formed into a cobra shape.The physician did not have complete visibility during the procedure, but states that the device did not sit properly so he resheathed the device twice in an attempt to fix the cobra shape, without success.The device was removed and deployed outside the patient, still taking on the cobra shape.Another physician resheathed the device and deployed it (outside the patient) and the occluder took on the proper shape.Tissue was noted on the device when initially removed from the patient.A new 24mm amplatzer septal occluder was used to successfully complete the procedure.The procedure was not significantly extended.The physician did see the deformation as a malfunction but reported it was not one that would result in patient harm.The patient was hemodynamically stable throughout the procedure and was discharged.
|
The reported event of deformity upon deployment could not be confirmed.The investigation confirmed the device met functional specifications when analyzed at abbott under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.Information from the field indicated that a 10f delivery sheath was used.Please note, per the instructions for use artmt100116885 ver a, the recommended delivery system size for this device is 9f.
|