Model Number 9-ASD-MF-030
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/24/2018
Event Type  Injury  
Event Description
On (b)(6) 2018, a 30mm amplatzer cribriform occluder was selected for use.During deployment, both left and right atrial discs were mushroom shaped.The device was resheathed and removed from the patient and the deformation was duplicated outside the body.A second 30mm amplatzer cribriform occluder was successfully implanted.The patient was hemodynamically stable throughout the procedure and the procedure was only extended by 5 minutes.The patient is reported stable and discharged.
 
Manufacturer Narrative
The reported event of "both left and right atrial discs were mushroom shaped" could not be confirmed.The investigation confirmed the device met all functional and dimensional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER CRIBRIFORM OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7879567
MDR Text Key120378697
Report Number2135147-2018-00154
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010304
UDI-Public00811806010304
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000039/S013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model Number9-ASD-MF-030
Device Catalogue Number9-ASD-MF-030
Device Lot Number6311458
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age65 YR
Patient Weight58
-
-