On (b)(6) 2018, a 30mm amplatzer patent foramen occluder was selected for use.The device's left atrial disc deployed in a balloon-like form within the patient.The device was resheathed and replaced with a 35mm amplatzer patent foramen occluder which was successfully implanted.The patient remained hemodynamically stable throughout the procedure and did not have any adverse health consequences.The procedure was not significantly extended.Patient's information (e.G.Age, weight, gender, ethnicity, race, patient's initials) cannot be handled by abbott in absence of patient's written consent as required by the personal data protection national legislation.National legislation prevents the recording of such information.A written consent has not been obtained in this case; therefore, this information is not available.
|