ST. JUDE MEDICAL, COSTA RICA LTDA TRIFECTA¿ GT VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number TFGT-23A
Device Problems Microbial Contamination of Device (2303); Obstruction of Flow (2423)
Patient Problems Corneal Pannus (1447); Endocarditis (1834)
Event Date 08/24/2018
Event Type  Injury  
Manufacturer Narrative
The reported event of endocarditis and pannus could not be confirmed.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
On (b)(6) 2016, a 23mm trifecta gt valve was implanted.On (b)(6) 2018, the valve was explanted due to valve endocarditis.The valve was observed sealed in.Pannus formation expanded from the aortic wall towards the struts and the polyester sewing ring were observed.Cloudy secretion was found in between and underneath the valve.Massive vegetation was found underneath the valve.The patient was reported stable.In the user's opinion, this was not a device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIFECTA¿ GT VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7879570
MDR Text Key120379766
Report Number3008452825-2018-00316
Device Sequence Number1
Product Code LWR
UDI-Device Identifier05415067018229
UDI-Public05415067018229
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100029/S021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/27/2017
Device Model NumberTFGT-23A
Device Catalogue NumberTFGT-23A
Device Lot Number5301016
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/27/2018
Initial Date FDA Received09/15/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-