Model Number 9-PFO-030
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/30/2018
Event Type  Injury  
Event Description
During the deployment of the 30 mm amplatzer pfo occluder, the left atrial disc remained elongated.This deformation was not attributed to patient anatomy.The device was not used and replaced with a second 30 mm amplatzer pfo occluder.The procedure was not significantly extended due to this event.The patient remained hemodynamically stable throughout the procedure and is reported to be stable and has been discharged.
 
Manufacturer Narrative
The reported event of deformity upon deployment could not be replicated.One photo of an echo was received from the field, which appeared to show the distal disc of an occluder deploy elongated; however, the device met functional specifications when analyzed at abbott under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PFO OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7884893
MDR Text Key120537221
Report Number2135147-2018-00156
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Model Number9-PFO-030
Device Catalogue Number9-PFO-030
Device Lot Number6487819
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/31/2018
Initial Date FDA Received09/18/2018
Supplement Dates Manufacturer Received11/29/2018
Supplement Dates FDA Received12/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-