• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION SWIFT-LOCK ANCHOR; SCS ANCHOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION SWIFT-LOCK ANCHOR; SCS ANCHOR Back to Search Results
Model Number 1192
Device Problem Fracture (1260)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/31/2018
Event Type  Injury  
Manufacturer Narrative
The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.The manufacturer defers to the patient's physician regarding medical history.
 
Event Description
It was reported that during the lead revision procedure (reference mfr report# 1627487-2018-09095) on (b)(6) 2018, the anchor was found to be damaged.As a result, physician explanted and replaced the anchor.Reportedly, effective therapy is confirmed post-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIFT-LOCK ANCHOR
Type of Device
SCS ANCHOR
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
seema wakdikar
6901 preston road
plano, TX 75024
9725268205
MDR Report Key7898772
MDR Text Key121042037
Report Number1627487-2018-08710
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067024084
UDI-Public05415067024084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P092371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/25/2020
Device Model Number1192
Device Lot Number6366468
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3189, SCS LEAD; MODEL 3772, SCS IPG
Patient Outcome(s) Other;
Patient Age53 YR
-
-