• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3357-40C
Device Problems Over-Sensing (1438); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/20/2018
Event Type  malfunction  
Event Description
It was reported that an asymptomatic patient presented in clinic during follow up.The cardiac resynchronization therapy defibrillator exhibited nonsustained right ventricular oversensing due to post-paced t-wave oversensing on the ventricular channel, observed on stored egm.The patient did not receive therapy during the oversensing.Programming changes were discussed.
 
Event Description
New information received indicated that the device was reprogrammed.On (b)(6) 2019, the patient presented in clinic and upon interrogation, post-paced t-wave oversensing was still noted.No programming changes were made due to the inability to perform defibrillation threshold testing.It was unknown why the defibrillation threshold testing could not be performed.Patient was stable and will continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7903706
MDR Text Key121571306
Report Number2017865-2018-14368
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508186
UDI-Public05414734508186
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2017
Device Model NumberCD3357-40C
Device Catalogue NumberCD3357-40C
Device Lot NumberA000014566
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2018
Initial Date FDA Received09/24/2018
Supplement Dates Manufacturer Received01/14/2019
Supplement Dates FDA Received01/25/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Weight94
-
-