Model Number EX1150
Device Problems Thermal Decomposition of Device (1071); Failure to Power Up (1476)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/11/2018
Event Type  malfunction  
Event Description
Hold for om 9/27/2018.It was reported that the merlin transmitter was no longer working due to a lightning strike.It was noted that the prongs blew off the plug in for the transmitter and charring was observed.The transmitter was replaced.No patient consequences were reported.
 
Manufacturer Narrative
Final analysis confirmed the reported event, however the alleged charring of the power adapter was not confirmed as this was missing and had been cut from the transmitter prior to receipt.There was no external physical damage to the transmitter.Internal inspection revealed physical damage to u-14 chips and multiple burn components to the main printed circuit board (pcb).In conclusion it was determined that the transmitter was hit by high current flow through the telephone line damaging the main pcb.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7908262
MDR Text Key121901761
Report Number2017865-2018-14438
Device Sequence Number1
Product Code DRG
UDI-Device Identifier05414734504799
UDI-Public05414734504799
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEX1150
Device Catalogue NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/11/2018
Initial Date FDA Received09/25/2018
Supplement Dates Manufacturer Received10/01/2018
Supplement Dates FDA Received10/15/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-