Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problem Unspecified Infection (1930)
Event Date 03/25/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: biolox delta ceramic liner # item 110003628 lot 3396322, taperloc stem # item 51-107120 lot 3442879, delta ceramic femoral head # item 650-0661 lot 2015111754, unknown screw.(b)(6).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2018-09040, 0001825034-2018-09042.Customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted in the patient.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the patient developed deep infection six weeks post initial surgery.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Udi - (b)(4).Reported event was unable to be confirmed due to limited information received from the customer.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient developed deep infection six weeks post initial surgery.Infection was resolved after i&d surgery and antibiotics treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 PPS LTD ACETABULAR SHELL
Type of Device
PROSTHESIS HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7911120
MDR Text Key121777899
Report Number0001825034-2018-09038
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 06/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number010000667
Device Lot Number3441507
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age41 YR
Patient Weight82
-
-