• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP SHOULDER; GLENOID, HEAD W/RETAINING SCREW, RSP, 32MM/-4MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP SHOULDER; GLENOID, HEAD W/RETAINING SCREW, RSP, 32MM/-4MM Back to Search Results
Model Number 508-32-103
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Discomfort (2330)
Event Date 09/07/2018
Event Type  Injury  
Event Description
Revision surgery - due to the patient experiencing discomfort and some dislocation of shoulder.
 
Manufacturer Narrative
The reason for this revision surgery was due to dislocation and discomfort.The previous surgery and the surgery detailed in this investigation occurred 4.5 years apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to dislocation and discomfort.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Type of Device
GLENOID, HEAD W/RETAINING SCREW, RSP, 32MM/-4MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7911934
MDR Text Key121842896
Report Number1644408-2018-00861
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912024747
UDI-Public(01)00888912024747
Combination Product (y/n)N
PMA/PMN Number
K051075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2020
Device Model Number508-32-103
Device Catalogue Number508-32-103
Device Lot Number864C1404
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/07/2018
Initial Date FDA Received09/26/2018
Supplement Dates Manufacturer Received11/01/2018
Supplement Dates FDA Received11/02/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
506-03-114, LOT 831C1163; 506-03-114, LOT 831C1163; 506-03-126, LOT 834C1064; 506-03-130, LOT 835C1120; 508-00-000, LOT 848C1408; 508-00-032, LOT 855C1271
Patient Outcome(s) Required Intervention;
Patient Age58 YR
-
-