U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC. FREESTYLE LIBRE; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC. FREESTYLE LIBRE; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71702-01
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Abscess (1690); Unspecified Infection (1930); Pain (1994)
Event Date 09/01/2018
Event Type  Injury  
Manufacturer Narrative
The customer's products have been requested for investigation.A follow-up report will be filed once the product is returned or additional information is obtained.The date of manufacture is unknown.The date listed is the date when abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported experiencing an infection while wearing an adc freestyle libre sensor.It was further reported that on (b)(6) 2018 customer noticed ¿pain, pus, a red ring and a hard lump¿ at the insertion site.Customer noted having contact with a healthcare provider on (b)(6)2018 and was diagnosed with an abscess.Customer was prescribed flucloxacillin and fusidic acid, (antibiotics).There was no report of death or permanent injury associated with this event.
 
Event Description
Customer reported experiencing an infection while wearing an adc freestyle libre sensor.It was further reported that on (b)(6) 2018 customer noticed ¿pain, pus, a red ring and a hard lump¿ at the insertion site.Customer noted having contact with a healthcare provider on (b)(6) 2018 and was diagnosed with an abscess.Customer was prescribed flucloxacillin and fusidic acid, (antibiotics).There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
Since no product has been returned, extended investigation has been performed for the reported complaint and determined that there was no indication that the product did not meet specification.Dhrs (device history review) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor kits.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.All review activities conducted above, including but not limited to the final release testing specifically associated with the manufacture of this product, are sufficient information in order to show if the product has met specifications prior to release.If the product is returned, the case will be re-opened and a physical investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC.
1360 south loop road
alameda CA 94502 7001
MDR Report Key7912444
MDR Text Key121846223
Report Number2954323-2018-07255
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71702-01
Was Device Available for Evaluation? No
Date Manufacturer Received06/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-