• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/28/2018
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction centrifuge bowl leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot g337 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot g337 for the reported issue shows no trends.Trends were reviewed for complaint categories, centrifuge bowl leak/break and system error.No trends were detected for each complaint category.Photographs were provided by the customer for evaluation.The photographs show the drive tube component of the kit intact with both drive tube bearings still connected to the drive tube.The outer centrifuge bowl had separated from the outer bowl cover.The base of the centrifuge bowl is intact and contained in the centrifuge bowl holder.A material trace of the bowl assembly and its components used to build lot g337 found no related nonconformances.A device history record review did not identify any related nonconformances and this kit lot had passed all lot release testing.The root cause of the centrifuge bowl break was most likely a separation of the outer bowl and outer bowl cover.The cause of the separation could not be determined based on the available information.No further action is required at this time.This investigation is now complete.(b)(4).
 
Event Description
The customer called to report a centrifuge bowl leak/break during the treatment procedure.The customer stated the break occurred during the purging air phase of the procedure.The customer stated a system error alarm was received after the bowl break occurred.The customer aborted the procedure and did not return blood to the patient.The customer has returned photographs for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS, INC.
bedminster NJ 07921
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key7915772
MDR Text Key123371880
Report Number2523595-2018-00145
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)G337(17)200601
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2020
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXUSA
Device Lot NumberG337
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/28/2018
Initial Date FDA Received09/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age23 YR
Patient Weight56
-
-