ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3361-40C
Device Problems Incorrect Interpretation of Signal (1543); Therapeutic or Diagnostic Output Failure (3023); Data Problem (3196)
Patient Problem Dizziness (2194)
Event Date 09/08/2018
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented in the emergency room after experiencing a shock from their implantable cardioverter defibrillator experiencing dizziness.Upon interrogation, the patient was in ventricular tachycardia (vt) and had received high voltage therapies.Later that evening, the patient was cardioverted with an external shock.Upon review, the egm for a vt episode was missing.On (b)(6) 2018, the physician elected to perform a defibrillation threshold test.During the test, there was an instance where the patient¿s induced vt rhythm was mistaken as noise, withholding therapy.Additionally, fluoroscopy indicated that the right ventricular lead exhibited an extrusion.Options were discussed and programming changes were made.The patient was doing well following the events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7918541
MDR Text Key122030971
Report Number2017865-2018-14576
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Model NumberCD3361-40C
Device Lot NumberS000018406
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/14/2018
Initial Date FDA Received09/28/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/04/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
(B)(4)
Patient Outcome(s) Required Intervention;
-
-