• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM/3.5MM VA-LCP POSTLAT DSTL HUM PL 7H/RT/127MM-LONG; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7MM/3.5MM VA-LCP POSTLAT DSTL HUM PL 7H/RT/127MM-LONG; PLATE, FIXATION, BONE Back to Search Results
Model Number 02.117.207
Device Problems Fitting Problem (2183); Device Damaged by Another Device (2915); Device-Device Incompatibility (2919)
Patient Problems No Consequences Or Impact To Patient (2199); No Code Available (3191)
Event Type  malfunction  
Manufacturer Narrative
Additional product code: hwc.Complainant device is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, the variable angle (va) self tapping locking screw went through the va locking compression plate (lcp) posterolateral distal humerus plate.The surgeon was using the va elbow system.The case was going fine until the last screw placement into the plate.The surgeon predrilled with a drill bit on a nominal angle.The surgeon powered the locking screw down, and stopped short of the plate and went to tighten by hand and the screw went directly through plate countersinking into the bone.The screw was not into the joint so the surgeon decided to leave the screw implanted in the bone.No action was done due to it would require taking all of the screws and the plate off and starting over.If the screw migrates into the joint, the patient would have to come back for a revision.It is unknown if there was a surgical delay.Patient and surgical outcome is unknown.Concomitant device reported: drill / drill bit (part# 323.062 lot# unknown, quantity# 1).This complaint involves one (1) device, 2.7mm/3.5mm va-lcp postlat dstl hum pl 7h/rt/127mm-long.This report is 2 of 2 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7MM/3.5MM VA-LCP POSTLAT DSTL HUM PL 7H/RT/127MM-LONG
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7920627
MDR Text Key122479155
Report Number2939274-2018-53995
Device Sequence Number1
Product Code HRS
UDI-Device Identifier10886982036162
UDI-Public(01)10886982036162
Combination Product (y/n)N
PMA/PMN Number
K120717
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number02.117.207
Device Catalogue Number02.117.207
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/04/2018
Initial Date FDA Received09/28/2018
Supplement Dates Manufacturer Received10/24/2018
Supplement Dates FDA Received10/24/2018
Patient Sequence Number1
Patient Age79 YR
-
-