• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) UNIFY QUADRA CRT-D, US MODEL, DF1COMP; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) UNIFY QUADRA CRT-D, US MODEL, DF1COMP; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3249-40
Device Problems Component Missing (2306); Connection Problem (2900)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/26/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that during the implant procedure, the left ventricular lead could not be fixed properly to the device resulting in ineffective pacing.It was noted that the screw under the silica gel was missing in the device.Device was not used and was replaced successfully.Patient¿s condition was stable.
 
Manufacturer Narrative
The reported event of missing set screw prior to implant could not be confirmed in the lab.While a set screw was missing and septum was heavily damaged when returned, it is undetermined as to when this occurred.Interrogation of the device revealed the device was above the elective replacement indicator when received.Pacing, sensing, impedance, hv output, patient notifier was tested and no anomaly was detected.The cause of the field event remains undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY QUADRA CRT-D, US MODEL, DF1COMP
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7927295
MDR Text Key122455945
Report Number2938836-2018-11053
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734504539
UDI-Public05414734504539
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Model NumberCD3249-40
Device Catalogue NumberCD3249-40
Device Lot NumberS000062071
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient Weight66
-
-