• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HE; BHR ACETABULAR CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD RESURFACING FEMORAL HE; BHR ACETABULAR CUP Back to Search Results
Catalog Number 74121150
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190); Appropriate Term/Code Not Available (3191)
Patient Problems Failure of Implant (1924); Pain (1994); Toxicity (2333); Injury (2348); Ambulation Difficulties (2544); Test Result (2695)
Event Date 08/27/2018
Event Type  Injury  
Event Description
It was reported that left hip revision surgery was performed due to pain, difficulty walking, elevated chromium and cobalt levels, metallosis and device failure.
 
Manufacturer Narrative
As of today, device return and additional information has been requested for this complaint but has not become available.Since neither the underlying medical documents nor device part details were received for investigation no thorough medical investigation, manufacturing record review and assessment of the reported event can be performed.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint.If the products or additional information become available in the future, this case will be reopened.Without additional information about this patient¿s particular case, our investigation remains inconclusive.No preventative or corrective action has been initiated as a result of this investigation.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.During the revision, the bhr head was removed.The bhr cup remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr head and cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.No further actions are required at this time.The available medical documents were reviewed.The reported elevated metal ion levels, pain, and granulomatosis-type changes noted intraoperatively are consistent with an adverse reaction to metal debris; however, without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the source cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.Also, without the implantation and pre-revision x-rays to determine initial implant anatomical placement and any micro-motion over time (for loosening), this cannot be ruled out as a contributory factor to the reported clinical reactions.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESURFACING FEMORAL HE
Type of Device
BHR ACETABULAR CUP
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa, warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key7929171
MDR Text Key122447152
Report Number3005975929-2018-00349
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup,Followup
Report Date 03/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2016
Device Catalogue Number74121150
Device Lot Number10MW31275
Was Device Available for Evaluation? No
Date Manufacturer Received09/10/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
FEMORAL HEAD, PART AND LOT # UNKNOWN.; FEMORAL HEAD, PART AND LOT # UNKNOWN
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
-
-