• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ESSENCE; COMPRESSOR, AIR, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD INNOSPIRE ESSENCE; COMPRESSOR, AIR, PORTABLE Back to Search Results
Model Number 1112266
Device Problem Degraded (1153)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/10/2018
Event Type  malfunction  
Event Description
It was reported that this device was getting warm and smoking.On return of the device for evaluation, no heat damage was observed inside or outside the device.However, it was observed that the mains cable was damaged to the extent that bare wires were visible, potentially exposing the patient to the risk of electrocution.No information has been provided as to how the mains cord came to be in this condition.The ifu for this device contains the warning 'the power supply cord cannot be replaced by the user.In case the power supply cord becomes damaged, contact philips respironics customer service for replacement (there is no service option)'.No patient harm or death has been reported and the device is not life sustaining/supporting.Internal reference ra (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNOSPIRE ESSENCE
Type of Device
COMPRESSOR, AIR, PORTABLE
Manufacturer (Section D)
RESPIRONICS RESPIRATORY DRUG DELIVERY (UK) LTD
chichester business park
city fields way, tangmere
chichester, west sussex PO20 2FT
UK  PO20 2FT
Manufacturer Contact
jayne newman
chichester business park
city fields way, tangmere
chichester, west sussex PO20 -2FT
UK   PO20 2FT
MDR Report Key7929179
MDR Text Key122461169
Report Number9681154-2018-00018
Device Sequence Number1
Product Code CAF
UDI-Device Identifier00383730001951
UDI-Public00383730001951
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K042655
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1112266
Device Catalogue NumberR1112261
Device Lot Number150625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Date Manufacturer Received09/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-