ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Use of Incorrect Control/Treatment Settings (1126)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/11/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that the cutter settings were changing on their own during setup for surgery.An alternate system was used to proceed with surgery.
 
Manufacturer Narrative
The company service representative examined the system and replicated the reported event.The pneumatic module was replaced to resolve the issue.The system was then tested and met all product specifications.The system was manufactured on september 9, 2014.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event can be attributed to non-conforming pneumatic module.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key7931309
MDR Text Key122656208
Report Number2028159-2018-02083
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/11/2018
Initial Date FDA Received10/03/2018
Supplement Dates Manufacturer Received12/04/2018
Supplement Dates FDA Received12/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-