The customer reports the nurse placed 110ml in the feeding bag of the connect feeding pump.The pump was programmed to deliver 96 ml to the patient over one hour.At the end of the hour, the pump indicated 96 ml was delivered, yet it visually appeared as though half or more of the feed was remaining in the bag.The nurse noted a kink in the tubing at the top of the cassette.It was confirmed that the tubing was kinked and the pump does not alert for occlusion.The pump was observed while running the test feed for a few minutes and the pump indicated that 10 ml was delivered but it was measured that only 2 ml was delivered.
|