• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - TIJUANA BAJA CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - TIJUANA BAJA CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE Back to Search Results
Catalog Number 101025
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/10/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available a supplemental report will be submitted.
 
Event Description
It was reported that during patient treatment with a cartridge bloodline, an external "reddish solution leakage" was observed at the level of the venous injection port.Treatment was interrupted and blood was returned to the patient.The cassette was replaced and treatment was resumed without further reported issues.The amount of external blood leak was not reported.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information was added.The actual device was not available for evaluation, however, eight retained samples were evaluated.Visual inspection and leak tests were performed and did not identify any abnormalities that could have contributed to the reported condition.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARTRIDGE
Type of Device
SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
Manufacturer (Section D)
BAXTER HEALTHCARE - TIJUANA BAJA
tijuana baja california
MDR Report Key7937870
MDR Text Key122807991
Report Number8030638-2018-00021
Device Sequence Number1
Product Code FJK
UDI-Device Identifier07332414007836
UDI-Public(01)07332414007836
Combination Product (y/n)N
PMA/PMN Number
K070414
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/15/2020
Device Catalogue Number101025
Device Lot Number1000185138
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/10/2018
Initial Date FDA Received10/05/2018
Supplement Dates Manufacturer Received10/09/2018
Supplement Dates FDA Received11/02/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age55 YR
-
-