The complained patient sample was tested on an abbott architect analyzer, resulting with the following values: tsh = 0.23 uiu/ml (reference range = 0.35 - 4.94 uiu/ml), ft4 = 1.64 ng/dl (reference range = 0.70- 1.48 ng/dl), and ft3 = > 20.00 pg/ml (reference range = 1.88-3.18 pg/ml).The ft3 and ft4 values measured with both roche and abbott analyzers were all above each assay's respective reference ranges.The differences in ft3 and ft4 values measured with the roche and abbott assays most likely relates to the overall setups of the assays, the antibodies used, differences in reference materials, and differences in the standardization methodology used.A general reagent issue can most likely be excluded.The investigation could not identify a product problem.The cause of the event could not be determined.
|